Product Reference Guide 5-1
Section 5
Calibration Procedures, Single Interval
NOTE
This section applies to scanner/scale models with single interv
al ONLY.
A number of situations require the scale to be calibrated. They are:
at initial installation of the scanner/scale
if the scale cannot be re-zeroed
if you change the weight measure from pounds to kilograms or
vice-v
ersa
if diagnostics indicate a calibration error
the weigh module has been replaced
Follow the procedures on the following pages to ensure that the scanner/
scale will meet W
eights and Measurement requirements.
LEGAL NOTE
Certification of the scanner/scale’s weighing apparatus is subject to
Federal, State and Local Weights and Measures statutes and is
restricted to authorized government agencies and/or duly registered
agents thereof. Anytime a scale is calibrated, it should be properly
sealed with a lead and wire or paper seal before being placed into ser
-
vice.
It is your responsibility to check with the appr
opriate authorities in your
area to ensure compliance with pertinent regulations before removing
any official seals or placing a newly calibrated scanner/scale into ser-
vice.